BINDER KBF P 720 constant climate chamber
The BINDER KBF P 720 constant climate chamber is a versatile and precise instrument that has been specially developed for use in laboratory and medical facilities. This climate chamber is ideal for the cultivation of microorganisms, bacteria and animal cultures as well as for various incubation and development processes. Thanks to its precise control of temperature and humidity, the KBF P 720 is ideal for observing and analyzing cell growth processes. Compared to similar products such as the BINDER KBWF 720 growth chamber, the KBF P 720 offers extended functionality and greater precision.
Technical data and functions
The BINDER KBF P 720 constant climate chamber offers a wide range of functions and technical specifications that make it an indispensable tool in any laboratory:
-
Temperature range without light cassettes: 0...70 °C
- Temperature rangewith 100 % illumination: 10...60 °C
-
Humidity range without light cassettes: 10...80 % r.h.
-
Humidity range with light cassettes: 10...75 % r.h.
-
Internal volume: 700 liters
-
Maximum load: 150 kg
-
Dimensions (W x H x D): 1250 x 1925 x 890 mm (outside), 973 x 1250 x 576 mm (inside)
-
Rated voltage: 200...230 V
-
Mains frequency: 50/60 Hz
-
Rated power: 3.5 kW
The KBF P 720 is equipped with a BINDER controller, which provides a clear display of all important parameters. The simple user interface and the graphical display options for the process parameters make operation much easier. The chamber also has a drift-free CO2 infrared absorption measuring system and an electronic fault auto-diagnosis system with visual and acoustic alarm.
Certifications and standards
The BINDER KBF P 720 constant climate chamber meets the requirements of DIN 12880:2007 and is equipped with standard-compliant hot air sterilization at 180 °C (DIN 58947). This ensures maximum safety and reliability in operation.
Application examples
The KBF P 720 is ideal for a wide range of applications in laboratory and medical facilities:
-
Cultivation of microorganisms: the oven provides optimum conditions for the growth and multiplication of bacteria and other microorganisms.
-
Cell cultures: The precise control of temperature and humidity enables the observation and analysis of cell growth processes.
-
Incubation and development processes: The KBF P 720 is ideal for carrying out incubation and development processes on various organisms.
-
Stability tests: The cabinet can be used for stability tests on pharmaceutical products and other materials.
FAQs
How is the temperature in the KBF P 720 controlled?
The temperature in the KBF P 720 is controlled by a precise temperature controller, which allows the temperature to be set and monitored accurately. The oven has a real-time clock and a timer, which allows flexible programming of the temperature cycles.
What safety functions does the KBF P 720 offer?
The KBF P 720 is equipped with an electronic fault auto-diagnosis system that triggers visual and audible alarms if a fault occurs. The oven also has a temperature selector monitor with visual and audible temperature alarm and a potential-free changeover contact for central monitoring.
Can the KBF P 720 be used for stability tests?
Yes, the KBF P 720 is ideal for stability tests. Thanks to its precise control of temperature and humidity, stability tests can be carried out under controlled conditions.
Conclusion
The BINDER KBF P 720 constant climate chamber is a powerful and versatile instrument that is indispensable in laboratory and medical facilities. With its precise control of temperature and humidity, its extensive safety functions and its simple operation, the KBF P 720 offers everything required for carrying out demanding experiments and tests. Similar products such as the BINDER KBF LQC 720 constant climate chamber also offer excellent performance, but with different specifications and functions.
Load max.: |
101 - 150 kg |
Width outside (incubators,...): |
1250 mm |
Width inside (incubators,...): |
973 mm |
Maximum temperature (heat): |
+ 100 °C |
Height outside (incubators,...): |
1925 mm |
Height inside (incubators,...): |
1250 mm |
Interior volume: |
601 - 700 l |
Cabinet type: |
constant climate chamber, climate chamber |
Depth outside (incubators,...): |
890 mm |
Depth inside (incubators,...): |
576 mm |
Lowest temperature (cooling): |
- 5 °C |
BINDER Constant climate chamber |
KBF P 720-230V1) |
KBF P 720UL-240V1) |
Performance Data Temperature |
|
|
Temperature range with 100% illumination [°C] |
10...60 |
10...60 |
Temperature range without illumination cassettes [°C] |
0...70 |
0...70 |
Max. heat compensation at 40 °C [W] |
1000 |
1000 |
Performance Data Climate |
|
|
Temperature range with 100% illumination [°C] |
20...60 |
20...60 |
Temperature range without illumination cassettes [°C] |
10...70 |
10...70 |
Temperature variation with illumination at 25 °C and 60 % RH [± K] |
1.2 |
1.2 |
Temperature variation with illumination at 40 °C and 75 % RH [± K] |
1.2 |
1.2 |
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] |
0.2 |
0.2 |
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] |
0.2 |
0.2 |
Humidity range with illumination cassettes, with illumination [% RH] |
10...75 |
10...75 |
Humidity range without illumination [% RH] |
10...80 |
10...80 |
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] |
2 |
2 |
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] |
2 |
2 |
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] |
4 |
4 |
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] |
5 |
5 |
Lightdata per Illumination Cassettes2) |
|
|
ICH compliant illumination for photo stability testing [lx] |
9000 |
9000 |
ICH compliant illumination for photo stability testing [W/m2] |
1.5 |
1.5 |
Electrical data |
|
|
Rated Voltage [V] |
200...230 |
200...240 |
Power frequency [Hz] |
50/60 |
50/60 |
Nominal power [kW] |
3.5 |
3.5 |
Unit fuse [A] |
16 |
16 |
Phase (Nominal voltage) [ph] |
1 |
1 |
Outer dimensions not incl. fittings and connections |
|
|
Width net [mm] |
1250 |
1250 |
Height net [mm] |
1925 |
1925 |
Depth net [mm] |
890 |
890 |
Wall clearance back [mm] |
100 |
100 |
Wall clearance sidewise [mm] |
200 |
200 |
Doors |
|
|
Inner doors |
2 |
2 |
Unit doors |
2 |
2 |
Internal Dimensions |
|
|
Width [mm] |
973 |
973 |
Height [mm] |
1250 |
1250 |
Depth [mm] |
576 |
576 |
Measures |
|
|
Interior volume [L] |
700 |
700 |
Net weight of the unit (empty) [kg] |
374 |
374 |
permitted load [kg] |
150 |
150 |
Load per rack [kg] |
45 |
45 |
Environment-specific data |
|
|
Energy consumption at 40 °C and 75 % RH [Wh/h] |
1850 |
1850 |
Sound-pressure level [dB(A)] |
53 |
53 |
Fixtures |
|
|
Number of shelves (std./max.) |
3/12 |
3/12 |
Number of illumination cassettes (std./max.) |
3/3 |
3/3 |
1) All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
2) Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde
There are no reviews yet.