The KBWF 720 from Binder is a very popular model that is suitable for every laboratory. You can use the model for a wide range of applications. In science laboratories or medical research institutes, these include in particular the cultivation of microorganisms, insect or animal cultures as well as various incubation or development processes. With the KBWF 720 you can also observe the behaviour of cells and their growth. This helps you to draw specific conclusions during the examination.
The KBWF 720: Reliable incubator for laboratory and medicine
The BINDER KBWF 720 growth chamber is often used in laboratories in biology, chemistry, medicine or the food industry. It allows you to individually control, for example, the temperature, humidity and other conditions such as the carbon dioxide and oxygen content for the formation and maintenance of a specific microclimate in the atmosphere of the incubator.
Depending on which model you choose, it has different features:
- Ergonomic control and information panel
- Handy handle bar
- Optional drawers and shelves
- Stainless steel interior
- Easy to clean
- Low power consumption
- Environmentally friendly thanks to low GWP value
- Low noise level
- Practical locking mechanism
- Membrane keypad with clear symbols
- Digital temperature display and setting via LED display
- Optimum temperature distribution in the entire usable space
- And many other special features!
Please contact our service team.
Cabinet type: climatic cabinet, growth cabinetMax. load: 101 - 150 kgInterior volume: 601 - 700 l
For more information about the features of the KBWF 720, please contact the MedSolut service team.
We are here for you!
Our team of experienced consultants wants you to get the best instruments for your field. Do you have questions about the various versions, their functions or their manufacturers?
We will be happy to advise you and can always be reached by telephone on +49 30 2096579 00 or by e-mail at info@medsolut.com. We look forward to getting in touch with you!
Load max.: |
101 - 150 kg |
Width outside (incubators,...): |
1250 mm |
Width inside (incubators,...): |
973 mm |
Maximum temperature (heat): |
+ 70 °C |
Height outside (incubators,...): |
1925 mm |
Height inside (incubators,...): |
1250 mm |
Interior volume: |
601 - 700 l |
Cabinet type: |
climate chamber, growth chamber |
Depth outside (incubators,...): |
890 mm |
Depth inside (incubators,...): |
576 mm |
Lowest temperature (cooling): |
0 °C |
Item weight: |
374,00 kg
|
BINDER Growth chamber |
KBWF 720-230V1) |
Performance Data Temperature |
|
Temperature range without illumination cassettes [°C] |
0...70 |
Temperature range with 100% illumination [°C] |
10...60 |
Performance Data Climate |
|
Temperature range with 100% illumination [°C] |
20...60 |
Temperature range without illumination cassettes [°C] |
10...70 |
Humidity range with illumination cassettes, with illumination [% RH] |
10...75 |
Humidity range without illumination [% RH] |
10...80 |
Lightdata per Illumination Cassettes2) |
|
Daylight tubes [W/m2] |
38 |
Daylight tubes [lx] |
13000 |
Fluora growth lamps [W/m2] |
36 |
Fluora growth lamps [lx] |
10500 |
Arabidopsis lamps [W/m2] |
43 |
Arabidopsis lamps [lx] |
14000 |
Electrical data |
|
Rated Voltage [V] |
200...240 |
Power frequency [Hz] |
50 |
Nominal power [kW] |
3.5 |
Unit fuse [A] |
16 |
Phase (Nominal voltage) [ph] |
1 |
Outer dimensions not incl. fittings and connections |
|
Width net [mm] |
1250 |
Height net [mm] |
1925 |
Depth net [mm] |
890 |
Wall clearance back [mm] |
100 |
Wall clearance sidewise [mm] |
200 |
Doors |
|
Inner doors |
2 |
Unit doors |
2 |
Internal Dimensions |
|
Width [mm] |
973 |
Height [mm] |
1250 |
Depth [mm] |
576 |
Measures |
|
Interior volume [L] |
700 |
Net weight of the unit (empty) [kg] |
374 |
permitted load [kg] |
150 |
Load per rack [kg] |
45 |
Environment-specific data |
|
Sound-pressure level [dB(A)] |
59 |
Fixtures |
|
Number of shelves (std./max.) |
3/12 |
Number of illumination cassettes (std./max.) |
3/3 |
1) All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
2) Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde