KNF LIQUIPORT diaphragm liquid pump NF 1.100
The KNF LIQUIPORT diaphragm liquid pump N F 1.100 is a high-performance diaphragm pump specially designed for use in demanding industrial and laboratory-based applications. This pump is characterized by its reliability, flexibility and energy efficiency, making it a preferred choice for professionals who require a robust and durable solution. Compared to other models such as the KNF LIQUIPORT diaphragm liquid pump NF 300, the NF 1.100 offers a balanced combination of performance and versatility.
Technical data and features
The KNF LIQUIPORT diaphragm liquid pump NF 1.100 offers a wide range of technical specifications that make it suitable for various applications:
-
Pump type: Diaphragm pump
-
Maximum flow rate: 1.3 l/min
-
Pump head material: PP (KT), PVDF (TT), PTFE (FT)
-
Maximum suction head: 3 mWS
- Permissiblepressure: 6.0 bar
- Permissibleviscosity of the medium: 150 cSt
-
Maximum power consumption: 15 W
-
Voltage: 24V DC
-
Maximum current consumption: 0.63 A
-
Noise level: < 40 dBA
-
Pump protection class: IP 65
- Protection classpower supply unit: IP 40
-
Protection class: III
These technical data make the pump ideal for applications that require high chemical resistance and reliability. Similar to the KNF chemical-resistant diaphragm dosing pump SIMDOS 10 FEM 1.10 RC-P, the NF 1.100 is designed for handling aggressive substances.
Areas of application
The KNF LIQUIPORT diaphragm liquid pump NF 1.100 is versatile and suitable for a wide range of industrial and laboratory applications:
-
Degassing: removal of dissolved gases from liquids.
-
Distillation processes: Support in the separation of substances by distillation.
-
Filtration processes: Conveying liquids through filter media.
-
Drying of moist substances: Acceleration of the drying process through efficient liquid conveying.
The pump is ideal for use with aggressive chemicals and solvents thanks to its chemical resistance and its ability to convey, evacuate and compress without distortion.
Certifications and standards
The pump complies with important industry standards and norms, including
-
DIN ISO 1629 and 1043.1: Standards for material classification.
-
Protection class III: High safety standards for electrical equipment.
-
IP 65: Protection against dust and water jets.
FAQs
Which media can be pumped with the KNF LIQUIPORT diaphragm liquid pump NF 1.100?
The pump is suitable for a wide range of media, including aggressive chemicals and solvents. It can pump media with a viscosity of up to 150 cSt.
How loud is the pump in operation?
The pump operates at a noise level of less than 40 dBA, making it one of the quietest pumps in its class.
What protection classes does the pump offer?
The pump has protection class IP 65, which means that it is protected against dust and water jets. The power supply unit has protection class IP 40.
Does the pump require a lot of maintenance?
No, the pump is low-maintenance and designed for continuous operation, making it a reliable solution for continuous applications.
Conclusion
The KNF LIQUIPORT diaphragm liquid pump NF 1.100 is a first-class choice for professionals who need a reliable, flexible and energy-efficient pump. With its high chemical resistance and ability to pump a wide range of media, it is ideal for demanding industrial and laboratory-based applications. Similar products such as the KNF LIQUIPORT diaphragm liquid pump NF 100 also offer excellent performance and may be a suitable alternative depending on specific requirements.
Max. Conveying capacity: |
1,3 l/min |
Model of pump: |
diaphragm pump |
Shipping weight: |
5,00 kg
|
Item weight: |
1,00 kg
|
|
NF 1.100 _ _ .18S |
NF 1.100 _ _ .18RC |
Pump materials 1)
|
KT (PP) TT (PVDF) FT (PTFE) |
KT (PP) TT (PVDF) FT (PTFE) |
Flow rate 2) (l/min) |
0.2 - 1.3 |
0.2 - 1.3 |
Suction head (mWG) |
3 |
3 |
Permissible pressure [Bar g] |
6.0 |
6.0 |
Permissible viscosity of medium [cSt] |
150 |
150 |
Max. power consumption DC [W] |
15 |
15 |
Pump DC voltage [V] |
24V DC |
24V DC |
Max. current consumption DC RMS 24 V [A] |
0.63 |
0.63 |
Max. short-term peak current [A] |
0.8 |
0.8 |
Power supply protection |
Electronic overload protection |
Electronic overload protection |
Pump protection |
Electronic overload protection |
Electronic overload protection |
Nominal mains voltage [V] |
100 - 240V AC +/- 10% |
100 - 240V AC +/- 10% |
Frequency [Hz] |
50-60 Hz |
50-60 Hz |
Max. power consumption AC 100 V / 115 V / 240 V [W] |
15 / 15 / 16 |
15 / 15 / 16 |
Weight of pump [kg] 3)
|
1.0 |
1.0 |
Permissible ambient temperature [°C] |
+5 to +40°C |
+5 to +40°C |
Permitted medium temperature |
+5 to +80°C |
+5 to +80°C |
Rated speed [U/min] |
3000 U/min |
3000 U/min |
Noise level [dBA] |
< 40 dBA |
< 40 dBA |
Pump protection type |
IP 65 |
IP 65 |
Power supply protection type |
IP 40 |
IP 40 |
Protection class |
III |
III |
Analogue input |
Signal range |
- |
0-10V |
Signal range optional4)
|
- |
0-20mA |
Input resistance [O] |
- |
133 kO at 0-10V 510 O at 0-20mA |
Electric strength [V] TTL |
- |
24 V DC |
Digital input |
Signal range |
- |
Pull Up at 24V |
Electric strength [V] TTL |
- |
24V DC |
Low level (ON) |
- |
< 0.8V = low |
High level (OFF) |
- |
> 2.0V = high |
Pull-up resistance |
- |
10 kO |
Resistance to ground |
- |
43 kO |
Digital output |
Electric strength Open Collector [V] TTL |
- |
24V DC |
Load capability Open Collector [mA] TTL |
- |
10mA |
1) according to DIN ISO 1629 and 1043.1
2) Flow rate at atmospheric pressure
3) The weight may differ slightly from the stated value, depending on the version
4) Optional on request
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde
There are no reviews yet.