Systec pass-through autoclaves D-Series 2D
The Systec pass-through autoclaves D-Series 2D are specially designed for use in laboratories that place the highest demands on sterilization. These autoclaves offer a compact, fast and energy-efficient solution for the sterilization of materials and devices. Thanks to their robust design and the wide range of available programs, they are ideal for a variety of applications in research and industry. Comparisons such as with the Systec vertical floor-standing autoclave VX show that the D-Series 2D particularly stands out due to its pass-through function and high flexibility in program selection.
Features and advantages
The Systec D-Series 2D pass-through autoclaves are characterized by a variety of technical features that make them an excellent choice for demanding laboratory environments:
-
Versatile sterilization programs: Depending on the model, up to 100 different sterilization programs are available, including a special program for destruction sterilization.
-
Robust construction: The housing and support frame are made of corrosion-resistant stainless steel, which ensures a long service life and high reliability.
-
Rapid heating: The autoclaves are equipped with a powerful heating element that enables rapid heating.
-
Safety features: A safety lock and automatic hose valve ensure safe operation.
-
Expandable functions: Optional features include exhaust air filtration, rapid re-cooling and extension of the temperature and pressure range.
Technical data
The technical data of the Systec pass-through autoclaves D-Series 2D vary depending on the model. Here are the most important specifications:
-
Design: Horizontal, pass-through autoclave
-
Maximum temperature: 140°C
-
Maximum pressure: 4.0 bar
-
Power: 9.0 kW
-
Chamber diameter: 400 mm or 500 mm
-
Chamber height or depth: 750 mm or 1000 mm
-
Chamber volume: 95 l, 150 l or 200 l
Certifications and standards
The Systec pass-through autoclaves D-Series 2D comply with all relevant standards and certifications for laboratory equipment. These include, among others:
-
DIN EN 285: This standard specifies the requirements for large sterilizers that work with saturated steam.
-
CE marking: The autoclaves are CE-certified and therefore meet the basic health and safety requirements of the EU.
Application examples
The Systec pass-through autoclaves D-Series 2D are versatile and suitable for a wide range of applications in laboratories and industry:
-
Sterilization of laboratory equipment: Ideal for the sterilization of glassware, metal instruments and plastic containers.
-
Sterilization of culture media: Suitable for the sterilization of culture media in microbiology and biotechnology.
-
Waste sterilization: With the special program for destruction sterilization, contaminated waste can be sterilized safely and efficiently.
FAQs
How many sterilization programs are available?
Depending on the model, up to 100 different sterilization programs are available, including a special program for destruction sterilization.
What safety features are integrated?
The autoclaves are equipped with a safety lock and an automatic hose valve to ensure safe operation.
Can the temperature and pressure range be extended?
Yes, an optional extension of the temperature and pressure range is available.
Is there a data backup option?
Yes, an optional SD card slot can be integrated for data backup.
Conclusion
The Systec D-Series 2D pass-through autoclaves offer a powerful and flexible solution for sterilization in laboratories. With a variety of programs, robust safety features and optional extensions, they are ideal for demanding applications in research and industry.
Design: |
horizontal, pass-through autoclave |
Pressure max. (autoclave): |
4.0 bar |
Diameter of vessel: |
400 mm, 500 mm |
Height or depth of vessel: |
750 mm, 1000 mm |
Power (autoclave): |
9.0 kW |
Temperature max. (autoclave): |
140°C |
Volume (autoclave): |
95 l, 150 l, 200 l |
Type |
Chamber dimensions Ø x depth (mm) |
Chamber volume (liters) total / nominal |
Dimensions wall opening WxH (mm) |
Dimensions clean room side WxH (mm) |
Dimensions service side incl. Wall frame WxH (mm) |
DX-90 2D |
400 x 750 |
98 / 95 |
820 x 670 |
920 x 780 |
1030 x 1570 |
DX-150 2D |
500 x 750 |
155 / 150 |
920 x 770 |
1020 x 880 |
1130 x 1670 |
DX-200 2D |
500 x 1000 |
205 / 200 |
920 x 770 |
1020 x 880 |
1130 x 1670 |
All types: heating power 9 kW, voltage 380 - 400 V, 50/60 Hz, 16 A.
Modifications to other voltage as well as single phase operation available on request.