Your safety cabinet for research, medicine and laboratory
The SUPPLY of the manufacturerL Type 90 from DÜPERTHAL SICHERHEITSTECHNIK is often in demand as a safety cabinet. Our cabinets are suitable for a wide range of processes and applications. These include, for example, the protected storage of all kinds of materials such as liquids, (gas) bottles, chemicals, solvents or acids. Depending on the type of safety cabinet, your cabinet will also be equipped with shelves or full extension drawers for optimum storage of the contents to be stored, or with drip trays.
Technical details of the SUPPLY of the manufacturerL Type 90
- Cabinet type: safety cabinet|barrel cabinet
The protection of personnel is the be-all and end-all at every workplace. All substances such as flammable liquids, chemicals or gas cylinders must therefore be stored under the safest possible conditions. The SUPPLY of the manufacturerL Type 90 from DÜPERTHAL SICHERHEITSTECHNIK is extremely well suited for this purpose.
Our customer advisor will explain even more about our devices, the different versions or our manufacturer!
You still have questions about the device?
Our team of first-class service staff will ensure that you receive the best items for your laboratory. Do you have questions about our wide range of safety cabinet devices, their features or the respective manufacturers?
We will be happy to assist you and are always available by phone at +49 30 2096579 00 or by e-mail at info@medsolut.com. We look forward to hearing from you!
Fire resistance: |
90 min |
Cabinet: |
safety storage cabinet, drum cabinet |
Doors, drawers: |
wing doors |
Item weight: |
873,00 kg
|
Düperthal Safety cabinet |
|
SUPPLY XXL Type 90 |
Cabinet W x D x H (mm) exterior |
|
1655 x 745 x 2080 |
Cabinet W x D x H (mm) interior |
Left Right |
744 x 595 x 15151) 744 x 595 x 11552)
|
Load bearing capacity with uniformly distributed load (maximum loadability/shelf (kg) |
Shelf Grating Heavy load drawer |
75 250 2503)
|
Maximum volume of the largest individual container which may be stored in the cabinet (l) |
Left Right |
60 200 |
Collection volume of the bottom tray (l) |
Left Right |
66 220 |
Volumetric flow rate for 10-fold air exchange (appr. m3/ h) |
|
20 |
Required extraction flow rates for filling and transferring (active storage*) (appr. m3/ h) |
|
> 42 |
Pressure drop in the cabinet (Pa) |
|
< 1 |
Weight (kg) |
|
873 |
1) In addition, the bottom tray with about 31 cm height
2) In addition, the bottom tray with about 67 cm height
3) Heavy load drawer with safety locking for 200-l drum available as accessory
*The continuously upcoming (24 h) exhaust air stream of the installation must be monitored during filling and transferring (active storage), to detect failure of the required air stream and to at least trigger an alarm.