KNF LIQUIPORT diaphragm liquid pump NF 1.300
The KNF LIQUIPORT diaphragm liquid pump NF 1.300 is a versatile and robust pump specially developed for use in demanding laboratory and industrial environments. This pump is characterized by its chemical-resistant nature and high efficiency, making it ideal for applications in biotechnology, medical technology and university research. Compared to similar products such as the KNF LIQUIPORT diaphragm liquid pump NF 1.100, the NF 1.300 offers a higher flow rate and greater flexibility.
Technical data and features
The KNF LIQUIPORT diaphragm liquid pump NF 1.300 offers a number of technical advantages that make it an excellent choice for various applications:
-
Pump type: Diaphragm pump
-
Max. Delivery rate: 3 l/min
-
Pump head material: PP, PVDF, PTFE
-
Maximum suction head: 3 mWS
- Permissiblepressure: 6.0 bar
- Permissibleviscosity of the medium: 150 cSt
-
Max. DC power consumption: 29 W
-
DC pump voltage: 24V DC
-
Max. Current consumption DC RMS: 1.2 A
-
Max. Short-term peak current: 1.3 A
-
Power supply unit and pump fuse: Electronic overload protection
-
Rated voltage power supply unit: 100-240V AC +/- 10%
-
Frequency: 50-60 Hz
-
Max. Power consumption AC: 30-32 W
-
Weight of the pump: 1.5 kg
- Permissibleambient temperature: +5 to +40°C
- Permissiblemedia temperature: +5 to +80°C
-
Nominal speed: 3000 rpm
-
Noise level: < 40 dBA
-
Pump protection class: IP 65
- Protection classpower supply unit: IP 40
-
Protection class: III
Areas of application
The KNF LIQUIPORT diaphragm liquid pump NF 1.300 is particularly suitable for:
-
Degassing: Removal of dissolved gases from liquids.
-
Distillation processes: Support in the separation of liquids by evaporation and condensation.
-
Filtration processes: Conveying liquids through filter media.
-
Drying of moist substances: Efficient removal of moisture from various materials.
This pump is ideal for use with aggressive chemicals and can be used in laboratories, industrial environments and research facilities. Similar to the KNF chemical-resistant diaphragm dosing pump SIMDOS 10 FEM 1.10 RC-P, the NF 1.300 offers high chemical resistance and reliability.
Certifications and standards
The KNF LIQUIPORT diaphragm liquid pump NF 1.300 complies with the following standards and certifications:
-
DIN ISO 1629 and 1043.1: Standards for material quality.
-
Protection class III: High safety and protection against electrical hazards.
-
IP 65: Complete protection against dust and protection against water jets from any angle.
FAQs
Which chemicals can be pumped with the KNF LIQUIPORT diaphragm liquid pump NF 1.300?
The pump is resistant to chemicals and can pump a wide range of aggressive chemicals, including acids, bases and solvents.
What is the maximum flow rate of the pump?
The maximum flow rate is 3 liters per minute.
Is the pump suitable for continuous operation?
Yes, the pump has a continuous running function and is designed for continuous operation.
What protective measures are integrated?
The pump and power supply unit are equipped with electronic overload protection to prevent damage caused by overloading.
Conclusion
The KNF LIQUIPORT diaphragm liquid pump NF 1.300 is a powerful and versatile pump that is characterized by its chemical resistance, high flow rate and reliability. It is ideal for demanding applications in laboratories and industrial environments and offers outstanding performance and durability. For more information and advice, please contact our team. See also similar products to KNF LIQUIPORT Diaphragm Liquid Pump NF 300 for more options.
Max. Conveying capacity: |
3 l/min |
Model of pump: |
diaphragm pump |
Shipping weight: |
5,00 kg
|
Item weight: |
1,50 kg
|
|
NF 1.300 _ _ .18S |
NF 1.300 _ _ .18RC |
Pump materials 1)
|
KT (PP) TT (PVDF) FT (PTFE) |
KT (PP) TT (PVDF) FT (PTFE) |
Flow rate 2) (l/min) |
0.5 - 3.0 |
0.5 - 3.0 |
Suction head (mWG) |
3 |
3 |
Permissible pressure [Bar g] |
6.0 |
6.0 |
Permissible viscosity of medium [cSt] |
150 |
150 |
Max. power consumption DC [W] |
29 |
29 |
Pump DC voltage [V] |
24V DC |
24V DC |
Max. current consumption DC RMS 24 V [A] |
1.2 |
1.2 |
Max. short-term peak current [A] |
1.3 |
1.3 |
Power supply protection |
Electronic overload protection |
Electronic overload protection |
Pump protection |
Electronic overload protection |
Electronic overload protection |
Nominal mains voltage [V] |
100 - 240V AC +/- 10% |
100 - 240V AC +/- 10% |
Frequency [Hz] |
50-60 Hz |
50-60 Hz |
Max. power consumption AC 100 V / 115 V / 240 V [W] |
30 / 30 / 32 |
30 / 30 / 32 |
Weight of pump [kg] 3)
|
1.5 |
1.5 |
Permissible ambient temperature [°C] |
+5 to +40°C |
+5 to +40°C |
Permitted medium temperature |
+5 to +80°C |
+5 to +80°C |
Rated speed [U/min] |
3000 U/min |
3000 U/min |
Noise level [dBA] |
< 40 dBA |
< 40 dBA |
Pump protection type |
IP 65 |
IP 65 |
Power supply protection type |
IP 40 |
IP 40 |
Protection class |
III |
III |
Analogue input |
Signal range |
- |
0-10V |
Signal range optional4)
|
- |
0-20mA |
Input resistance [O] |
- |
133 kO at 0-10V 510 O at 0-20mA |
Electric strength [V] TTL |
- |
24 V DC |
Digital input |
Signal range |
- |
Pull Up at 24V |
Electric strength [V] TTL |
- |
24V DC |
Low level (ON) |
- |
< 0.8V = low |
High level (OFF) |
- |
> 2.0V = high |
Pull-up resistance |
- |
10 kO |
Resistance to ground |
- |
43 kO |
Digital output |
Electric strength Open Collector [V] TTL |
- |
24V DC |
Load capability Open Collector [mA] TTL |
- |
10mA |
1) according to DIN ISO 1629 and 1043.1
2) Flow rate at atmospheric pressure
3) The weight may differ slightly from the stated value, depending on the version
4) Optional on request
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde
There are no reviews yet.