ENVAIR safety cabinet eco safe Comfort Class II
The ENVAIR eco safe Comfort Class II safety workbench is a state-of-the-art safety workbench specially developed for applications in optics, laser, semiconductor and electronics technology. This workbench provides a sterile and dust-free environment that ensures the protection of both employees and products. Thanks to its outstanding features and certifications, it is a preferred choice for professionals who need to comply with the highest safety and quality standards. Similar to the FASTER SafeFAST Light safety cabinet, it offers a reliable solution for demanding work environments.
Technical data and features
The ENVAIR eco safe Comfort Class II safety cabinet is characterized by the following technical data and features:
-
Safety class: Class II
-
Width: 1001 mm - 1100 mm, 1301 mm - 1400 mm, 1601 mm - 1700 mm, 1901 mm - 2000 mm
-
Depth: 801 mm - 900 mm
-
Height: 1401 mm - 1500 mm
-
General dimensions (WxHxD): 1045x1500x860 mm to 1960x1500x860 mm
-
Usable dimensions (WxHxD): 887x740x580 mm to 1802x740x580 mm
-
Working opening: 200 mm
-
Max. Front opening: 440 mm
-
Weight: 170 kg to 260 kg
-
Exhaust air: 290 m³/h to 585 m³/h
-
Noise level: < 53 dB(A) to < 56 dB(A)
-
Light intensity: >1100 lux to >1300 lux
-
Electrical connection: 1Ph+E - 230V, 50Hz
-
Power consumption: 281 W to 691 W
-
Total current: 1.9 A to 3.6 A
-
Protection class / IP: 1 / 20
-
Internal fuse protection: 6 A
-
Heat output: 175 W to 360 W
Certifications and standards
The ENVAIR safety cabinet eco safe Comfort Class II complies with the European standards EN 12469:2000 and is TÜV-tested. These certifications guarantee the highest safety standards and reliability in operation. In contrast to the FASTER safety cabinet SafeFAST Classic double blower, it offers comprehensive certification, which is essential for many applications.
Application examples
This safety cabinet is frequently used in the following areas:
-
Optics technology: For the manufacture and testing of optical components.
-
Laser technology: For the development and maintenance of laser systems.
-
Semiconductor technology: For the production and analysis of semiconductor components.
-
Electronics technology: For the assembly and testing of electronic circuits and devices.
FAQs
What advantages does the ENVAIR eco safe Comfort Class II safety cabinet offer?
-
Maximum safety standard: Protection for employees and products.
-
High-performance production: High reliability and durability.
-
Air testing: Ensures a sterile working environment.
-
Modular design: Flexibility and space saving.
-
Easy to use: User-friendly design and easy maintenance.
How often should the safety cabinet be serviced?
Regular maintenance is crucial for the safety and longevity of the workbench. It is recommended that the workbench is inspected at least once a year by qualified technicians. A maintenance contract can help to ensure regular inspections.
What types of filters are used and how often do they need to be changed?
The workbench uses integrated HEPA filters, which should be changed regularly depending on the frequency of use and the environment. Easy filter replacement is guaranteed to keep operating costs low.
Conclusion
The ENVAIR eco safe Comfort Class II safety cabinet is a first-class choice for professionals who require a sterile and safe working environment. With its extensive certifications, flexible application options and user-friendly features, it offers high added value for various industries. Similar to the ESCO NordicSafe - Microbiological Safety Cabinets Class II, it is a reliable solution for demanding applications.
| Item information |
Value |
| Width: |
1001 mm - 1100 mm, 1301 mm - 1400 mm, 1601 mm - 1700 mm, 1901 mm - 2000 mm |
| Device type: |
safety cabinet |
| Height: |
1401 mm - 1500 mm |
| Safety class: |
class II |
| Depth: |
801 mm - 900 mm |
| Safety cabinets |
ENVAIR eco safe Comfort Class II |
|
0.9 m |
1.2 m |
1.5 m |
1.8 m |
Overall Dimensions WxHxD (1) (mm) |
1045x1500x860 |
1350x1500x860 |
1655x1500x860 |
1960x1500x860 |
Useful Dimensions WxHxD (mm) |
887x740x580 |
1192x740x580 |
1497x740x580 |
1802x740x580 |
| Working aperture |
200 mm |
200 mm |
200 mm |
200 mm |
| Max front aperture |
440 mm |
440 mm |
440 mm |
440 mm |
| Weight |
170 kg |
196 kg |
225 kg |
260 kg |
| Exhaust flow rate |
290 m3/h |
390 m3/h |
485 m3/h |
585 m3/h |
| Noise level (2) |
< 53 dB(A) |
< 54 dB(A) |
< 55 dB(A) |
< 56 dB(A) |
| Lighting level |
>1100 lux |
>1200 lux |
>1300 lux |
>1300 lux |
| Electrical Data |
1Ph+E - 230V 50Hz |
1Ph+E - 230V 50Hz |
1Ph+E - 230V 50Hz |
1Ph+E - 230V 50Hz |
| Power consumption |
281 W |
390 W |
587 W |
691 W |
| Current consumption (2) (3) |
1.9 A |
2.1 A |
3.4 A |
3.6 A |
| Electrical class / IP |
1 / 20 |
1 / 20 |
1 / 20 |
1 / 20 |
| Internal electrical outlet |
6 A |
6 A |
6 A |
6 A |
| Heat emission |
175 W |
240 W |
295 W |
360 W |
(1) The total depth of the cabinet can be easily reduced to 780 mm by removing the back panel
(2) At operation condition according to EN12469: 2000
(3) Clean filters, lighting activated, internal outlet load excluded
Envair offers service in pure culture. We do not simply sell our units to you but will assist you gladly from A to Z.
Delivery:
We deliver all units ex works.
Set-up service and packaging disposal are included upon request.
Service:
- 24h at site service and 24h free of charge phone hotline
- 7 days a week
Spare parts guaranty:
- 10 years
- Availability within 24h
- Ex stock Emmendingen
Qualification:
- IQ/OQ
- GMP
- GAMP
- 21 CFR Part 11
Decontamination:
We carry out microbiological decontaminations with most modern H2O2 generators in Cleanrooms, Isolators and Safety Cabinets.
Maintenance:
- Microbiological Safety Cabinets DIN 12469
- Cytotoxic Dispensing Cabinets with KI-discus test DIN 12980
- Laminar Flow Cabinets
- Laboratory Fume Hoods
- Cytotoxic Dispensing Isolator DIN 12980
- Isolator EN 14644-7
- Cleanrooms
- Autoclaves
- CO2 Incubator
- Ultra Low Temperature Freezer -86 °C
Service by professional staff. All works are carried out by service technicians qualified by TÜV Nord.
Maintenance contract:
Regular terms of maintenance increase your safety.
With a maintenance contract we guarantee the complete check of all functions of your units. We will contact you reliably in regular terms.
Guaranty and Warranty:
Combined with an ENVAIR maintenance contract we offer a 2-year warranty to all units. Excepted are consumables and wearing parts as well as compressed air components.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde