The ELGA PURELAB flex 2 from VWS Germany is a state-of-the-art ultrapure water system that has been specially developed for the precise and sustainable production of pure or demineralized water. This system is ideal for laboratories, doctors' surgeries, hospitals and industrial applications. It offers reliable performance and is particularly energy efficient, making it a preferred choice over similar products on the market. Similar to the ELGA PURELAB flex 4, the flex 2 offers high water quality and ease of use.
Technical data and features
The ELGA PURELAB flex 2 is characterized by the following technical data and features:
-
Water treatment type: Ultrapure water system
-
Capacity: 2 l/min
-
Modules: UV module
-
Water source: from reverse osmosis system, from ion exchanger
-
Daily water volume: > 10 liters
-
Max. Flow rate: up to 2 l/min
-
Inorganic substances at 25°C: 18.2 MOhm-cm
-
Organic substances (TOC) - typical: < 5 ppb
-
Bacteria - typical (with final filter): < 1 CFU/10 ml
-
Bacteria - typical (with biofilter): < 1 CFU/10 ml
-
Endotoxin (with biofilter): < 0.001 EU/ml
-
DNase (with biofilter): < 20 pg/ml
-
RNase (with biofilter): < 0.002 ng/ml
-
Dimensions: Width 236 mm, depth 374 mm, min. height 900 mm, max. height 1020 mm
-
Operating weight: 11 kg
-
Installation: table/wall
-
Mains power consumption: 100 - 240 V ac, 50 - 60 Hz
-
System voltage (without pumps and UV): 24 V dc
-
Energy consumption at peak load: 120 W
-
Noise level: < 40 dBa
Areas of application
The ELGA PURELAB flex 2 can be used in a variety of ways and is ideal for
-
Laboratories: For the preparation of reagents and for chemical analysis.
-
Medical practices and hospitals: For the production of medicines and injection fluids.
-
Industry: For the preparation of tissue and cell cultures and for the production of ointments.
-
Car repair shops: For special cleaning processes.
Certifications and standards
The ELGA PURELAB flex 2 complies with all relevant certifications and standards required for use in medical and industrial environments. This ensures high quality and safety in water treatment.
FAQs
How does the UV module in the ELGA PURELAB flex 2 work?
The UV module in the ELGA PURELAB flex 2 uses UV light to kill microorganisms in the water. This provides additional purification and improves the water quality.
Which water sources can be used for the ELGA PURELAB flex 2?
The ELGA PURELAB flex 2 can draw water from reverse osmosis systems and ion exchangers. This enables flexible integration into existing water treatment systems.
How high is the energy consumption of the ELGA PURELAB flex 2?
The energy consumption at peak load is 120 W, which makes the ELGA PURELAB flex 2 an energy-efficient device.
Can the ELGA PURELAB flex 2 be mounted on the wall?
Yes, the ELGA PURELAB flex 2 can be installed both on a table and on the wall, allowing flexible placement.
Conclusion
The ELGA PURELAB flex 2 is a powerful and versatile ultrapure water system that is ideal for use in laboratories, medical facilities and industry. With its high water quality, energy efficiency and ease of use, it is an excellent choice for anyone who needs reliable ultrapure water. Similar to the ELGA PURELAB Classic UV, the flex 2 offers outstanding performance and quality.
Booster pump: |
without |
Power: |
2 l/min |
Modules: |
UV-modules |
Storage tank - tank: |
without |
Water treatment type: |
ultrapure water system |
Water source: |
from reverse osmosis system, of ion exchanger |
Shipping weight: |
16,00 kg
|
Item weight: |
11,00 kg
|
model |
PURELAB flex 2 |
Specifications Product water |
|
Daily water quantity |
> 10 liters |
Max. Flow rate |
up to 2 l/min |
inorganic substances at 25°C |
18,2 MOhm-cm |
organic substances (TOC) - typical |
< 5ppb 1)
|
Directly from the integrated tank |
- |
Bacteria - typical (with final filter) |
< 1 KBE/10 ml |
Bacteria - typical (with biofilters) |
< 1 KBE/10 ml |
Endotoxin (with biofilters) |
< 0,001 EU/ml |
DNase (with biofilters) |
< 20 pg/ml |
RNase (with biofilters) |
< 0,002 ng/ml |
Dimensions and weights |
|
dimensions |
width 236 mm depth 374 mm min. height 900 mm max. height 1020 mm |
operating weight |
11 kg (24,2 lb) |
installation |
Desk / Wall |
Specifications feedwater |
|
source of supply |
Drinking water with subsequent treatment, preferably by reverse osmosis (RO) or ion exchanger (SDI) - filtered or distilled |
conductivity |
< 1 µS/cm 2) |
ingredients |
|
hardness |
- |
Free chlorine |
< 0,05 ppm Cl2
|
chloramine |
< 0,05 ppm Cl2
|
chlorine Total |
< 0,05 ppm Cl2 |
silicate |
< 2 ppm |
carbon dioxide maximum |
< 0,1 ppm |
Fouling Index |
< 1 |
Iron / manganese |
- |
organic substances (TOC) |
- |
particle |
A 0.2 micron membrane prefilters is for all feedwater systems (except reverse osmosis) is recommended for longer tool life of final filters |
temperature |
4 - 40°C (recommended 10 - 25°C) |
flow (at 15°C) |
< 2 l/min |
Runoff Specification (free expiry) |
not required |
Feed water pressure |
|
maximum |
22 psi (1,5 bar) |
minimum |
minimum positive admission pressure |
Electrical specifications |
|
Mains power input |
100 - 240 V ac, 50 - 60 Hz |
System voltage (without pump and UV) |
24 V dc |
Energy consumption at peak load |
120 W |
noise level |
< 40 dBa |
1) Depending on the feedwater
2) A pretreatment cartridge at a feed water quality 1 µS / cm - 30 µS / cm necessary.