The Carbolite CR 250 °C cleanroom drying oven is a versatile and powerful appliance designed specifically for use in cleanrooms and laboratories. With a maximum temperature of 250 °C and excellent temperature homogeneity, this drying oven is ideal for a variety of applications, including drying, moisture extraction and cleanroom applications. Compared to similar products such as the Carbolite PF air circulation drying oven up to 300 °C, the CR 250 °C offers a specialized solution for cleanroom conditions.
Technical data and features
The Carbolite CR 250 °C cleanroom drying oven is characterized by the following technical data and features:
-
Oven type: Drying oven, heating oven
-
Maximum temperature: 250 °C
-
Temperature stability: ±0.2 °C
-
Temperature homogeneity: ±3.0 to ±5.0 °C at 250 °C
-
Internal volume: 30 to 1810 liters, depending on model
-
Max. load: 11 - 50 kg, 51 - 100 kg
-
Air flow: Forced
-
Dimensions (inside): Varies depending on model, e.g. 310 x 310 x 310 mm to 1220 x 1220 x 1220 mm
-
Dimensions (outside): Varies depending on model, e.g. 685 x 460 x 670 mm to 1930 x 1750 x 1580 mm
-
Maximum power: 1 kW to 18 kW, depending on model
-
Safety standard: ISO 14644-1 class 5
Certifications and standards
The Carbolite CR 250 °C clean room drying oven complies with the BS EN 61010-2-010:2003 and ISO 14644-1 Class 5 standards. These certifications ensure that the drying oven meets the highest safety and quality standards and is therefore ideal for use in sensitive environments such as clean rooms.
Application examples
The Carbolite CR 250 °C clean room drying oven is versatile and is particularly suitable for the following applications:
-
Drying and moisture extraction: ideal for removing moisture from samples and materials.
-
Clean room applications: Perfect for use in cleanrooms where a controlled environment is required.
-
Cultivation of microorganisms: Supports the cultivation and growth of microorganisms under controlled conditions.
-
Incubation and development processes: Suitable for carrying out incubation and development processes in laboratory practice.
FAQs
What power supply is required for the Carbolite CR 250 °C cleanroom drying oven?
The power supply varies depending on the model. Some models require a single-phase 220-240 V power supply, while other models require a three-phase 380-415 V power supply. It is important to specify the required power supply and mains connection when ordering.
What additional options and accessories are available for the Carbolite CR 250 °C cleanroom drying oven?
The CR 250 °C offers numerous customization options and accessories, including
- Over-temperature protection
- Special shelves and pull-outs
- Digital timers
- State-of-the-art temperature controllers
How long does it take to heat up the Carbolite CR 250 °C cleanroom drying oven?
The heat-up time varies depending on the model and is usually between 35 and 80 minutes to reach the maximum temperature of 250 °C.
Conclusion
The Carbolite CR 250 °C clean room drying oven is a powerful and versatile appliance that is ideal for use in clean rooms and laboratories. With its high temperature homogeneity, extensive customization options and stringent safety certifications, the CR 250 °C offers a reliable solution for a wide range of applications. Similar products such as the Carbolite Economy circulating air drying oven AX 250 °C also offer interesting alternatives for specific requirements.
Load max.: |
11 - 50 kg, 51 - 100 kg |
Width outside (incubators,...): |
460 mm, 620 mm, 940 mm, 1395 mm, 1750 mm |
Width inside (incubators,...): |
310 mm, 470 mm, 610 mm, 915 mm, 1220 mm |
Maximum temperature (heat): |
+ 250 °C |
Height outside (incubators,...): |
685 mm, 925 mm, 1145 mm, 1360 mm, 1670 mm, 1930 mm, 2235 mm |
Height inside (incubators,...): |
310 mm, 550 mm, 610 mm, 770 mm, 915 mm, 1220 mm, 1525 mm |
Interior volume: |
up to 50 l, 51 - 100 l, 101 - 200 l, 201 - 300 l, 301 - 400 l, 401 - 500 l, 801 - 900 l, 1501 - 2000 l |
Maximum power: |
1 kW, 1.5 kW, 2 kW, 2.5 kW, 3 kW, 4.5 kW, 6 kW, 12 kW, 18 kW |
Air convestion (incubators, ...): |
forced |
Suitable for: |
Clean room applications, Drying / moisture extraction |
Cabinet type: |
heating oven, Heating cabinet |
Safety standard: |
ISO 14644-1 Class 5 |
Depth outside (incubators,...): |
670 mm, 820 mm, 970 mm, 1580 mm |
Depth inside (incubators,...): |
310 mm, 470 mm, 610 mm, 1220 mm |
Carbolite
Cleanroom-
Drying cabinet
CR |
Maximum
temperature
(°C) |
Temperature-
Stability
(°C) |
Temperature-
Homogeneity
(°C) |
Heating time
(minutes) |
Recovery-
Time
(minutes) |
Dimensions
Inside
H x W x D
(mm) |
Dimensions
Outside
H x W x D
(mm) |
Grille
inclusive/
maximum |
Maximum load
per grille/
total
(kg) |
Chamber-
Volume
(litres) |
Maximum
power
(W) |
CR/30 |
250 |
±0,2 |
±3.0 at 250 °C |
35 |
4 |
310 x 310 x 310 |
685 x 460 x 670
(table-top unit*) |
2/2 |
10/20 |
30 |
1000 |
CR/70 |
250 |
±0,2 |
±3.0 at 250 °C |
35 |
4 |
310 x 470 x 470 |
685 x 620 x 820
(table-top unit*) |
2/2 |
10/20 |
68 |
1500 |
CR/130 |
250 |
±0,2 |
±4.0 at 250 °C |
35 |
4 |
550 x 470 x 470 |
925 x 620 x 820
(table-top unit*) |
3/5 |
10/40 |
121 |
2000 |
CR/180 |
250 |
±0,2 |
±5.0 at 250 °C |
58 |
5 |
770 x 470 x 470 |
1145 x 620 x 820
(table-top unit*) |
3/7 |
10/50 |
170 |
2500 |
CR/220 |
250 |
±0,2 |
±5.0 at 250 °C |
75 |
4 |
610 x 610 x 610 |
1360 x 940 x 970
(table-top unit*) |
3/5 |
15/45 |
227 |
3000 |
CR/330 |
250 |
±0,2 |
±5.0 at 250 °C |
80 |
6 |
915 x 610 x 610 |
1670 x 940 x 970
(floor standing unit*) |
4/8 |
15/60 |
340 |
4500 |
CR/450 |
250 |
±0,3 |
±5.0 at 250 °C |
75 |
9 |
1220 x 610 x 610 |
1930 x 940 x 970
(floor-standing unit*) |
5/11 |
15/75 |
450 |
6000 |
CR/840 |
250 |
±0,3 |
±5.0 at 250 °C |
- |
- |
1525 x 915 x 610 |
2235 x 1395 x 970
(floor standing unit) |
6 |
15/- |
850 |
12000 |
CR/1790 |
250 |
±0,3 |
±5.0 at 250 °C |
- |
- |
1220 x 1220 x 1220 |
1930 x 1750 x 1580
(floor standing unit) |
5 |
15/- |
1810 |
18000 |
Please note:
The minimum working temperature is about 30 °C above room temperature.
Temperature homogeneity measured in empty and closed chamber after a stabilisation phase
Maximum power and heating time with 240 V connection
Maximum load on the grilles possible with evenly distributed weight
The zone of homogeneous temperature is smaller than the total volume
* Optional with base
Power supply |
Volt |
Phase |
Amps per phase |
CR / 330 litres |
220-240 |
1 phase |
19,5 |
|
380-415 |
3 phases + N |
8,9 |
CR / 450 litres |
220-240 |
3 phase delta |
15,5 |
|
380-415 |
3 phases + N |
8,9 |
CR larger than 450 litres on request |
|
|
|
All models not listed here in the table can be operated with a 220-240 V, 16 A single-phase connection.
Please specify the desired power supply and mains connection when ordering!
Other power connection configurations on request.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde
There are no reviews yet.