The KBW 240 from Binder is a climate chamber and can be used in many areas in laboratories or medicine. The incubators and incubators from Binder are suitable for a range of processes and applications. These are, for example, in laboratories or medical research centres, the cultivation of microorganisms, bacteria or animal cultures or numerous incubation or development processes. The KBW 240 can also be used for detailed investigations of cell behaviour and growth.
Your incubator for research, medicine and laboratory work
Binder incubators are mainly used by our customers for both bacterial and eukaryotic cells. To create and maintain a specific microclimate, the temperature, humidity and other conditions such as the carbon dioxide and oxygen content in the incubator atmosphere can be regulated.
Depending on the design, the incubator has different functions:
- Ergonomic control and information panel
- Handy handle bar
- Optional drawers and shelves
- Stainless steel interior
- Easy to clean
- Low power consumption
- Environmentally friendly thanks to low GWP value
- Low noise level
- Practical locking mechanism
- Membrane keypad with clear symbols
- Digital temperature display and setting via LED display
- Optimum temperature distribution in the entire usable space
- And many other special features!
Please contact our service team.
Cabinet type: Climatic cabinetMax. load: 51 - 100 kgInterior volume: 201 - 300 l
Our trained service team will be happy to provide you with more detailed information on equipment features.
Get advice from the MedSolut team now!
Our expert team of advisors aims to ensure that you receive the right products for your work area. Do you have questions about our various models, their functions or our manufacturers?
We look forward to hearing from you and will be happy to assist you! You can reach us by phone at +49 30 2096579 00 or by e-mail at info@medsolut.com.
Load max.: |
51 - 100 kg |
Width outside (incubators,...): |
930 mm |
Width inside (incubators,...): |
650 mm |
Maximum temperature (heat): |
+ 70 °C |
Height outside (incubators,...): |
1460 mm |
Height inside (incubators,...): |
785 mm |
Interior volume: |
201 - 300 l |
Cabinet type: |
climate chamber |
Depth outside (incubators,...): |
880 mm |
Depth inside (incubators,...): |
485 mm |
Lowest temperature (cooling): |
0 °C |
BINDER Climate chamber |
KBW 240 (E5.1) |
Exterior dimensions |
|
Width (mm) |
930 |
Height (incl. casters) (mm) |
1460 |
Depth (incl. door handle, I-triangle, connection 80 mm) (mm) |
880 |
Wall clearance, rear (mm) |
100 |
Wall clearance, side (mm) |
100 |
Steam space volume (l) |
348 |
Number of doors (ea.) |
1 |
Inner glass door(s) (ea.) |
1 |
Interior dimensions |
|
Width (mm) |
650 |
Height (mm) |
785 |
Depth (mm) |
485 |
Interior volume (l) |
247 |
Racks (number standard/max.) |
2 / 9 |
Load per rack (kg) |
30 |
Permitted total load (kg) |
100 |
Weight (empty) (kg) |
202 |
Variable position illumination cassettes |
2 |
Temperature data (without illumination) |
|
Temperature range (°C) |
0 - 70 |
Temperature variation (± K) |
0,5 |
Temperature fluctuation (± K) |
0,1 |
Max. heat compensation up to 40 °C (W) |
350 |
Temperature data (with 100 % illumination) |
|
Temperature range (°C) |
5 - 60 |
Temperature variation (± K) |
0,5 |
Temperature fluctuation (± K) |
0,1 |
Max. heat compensation up to 40 °C (W) |
250 |
Illumination data (per illumination cassette) |
|
Daylight tubes (Lux) / (W/m2) |
9.000 / 24 |
Fluora® growth lamps (Lux) / (W/m2) |
7.500 / 23 |
Arabidopsis tubes (Lux) / (W/m2) |
11.000 / 32 |
Electrical data |
|
IP protection class acc. to EN 60529 |
IP 20 |
Voltage (± 10 %) 50 / 60 Hz (V) |
200 - 240, 1 N ~ |
Nominal power (kW) |
1,4 |
Energy consumption 1) without illumination at 0 °C (W) with illumination at 4 °C (W) with illumination at 25 °C (W) with illumination at 37 °C (W) |
360 520 495 525 |
Noise level (dB (A)) |
52 |
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.
Kunde