BINDER KBW 720 growth chamber
The BINDER KBW 720 growth chamber is a versatile all-rounder that has been specially developed for use in research, medicine and laboratory work. This growth chamber offers optimum conditions for the cultivation of microorganisms, bacteria and animal cultures as well as for various incubation and development processes. Thanks to its precise temperature and humidity control, the KBW 720 is ideal for observing cell behavior and growth, allowing you to draw well-founded conclusions in your investigations. Similar to the BINDER KBF-S 720 constant climate chamber, the KBW 720 provides a reliable and uniform environment for your experiments.
Technical data and equipment
The BINDER KBW 720 growth chamber is characterized by a number of technical features that make it an indispensable instrument in any laboratory:
-
Temperature range: 0 °C to +70 °C without light cassettes, 5 °C to +60 °C with 100 % illumination
-
Internal volume: 698 liters
-
Maximum load: 150 kg
-
Dimensions:
- Inside: 970 mm (width) x 1250 mm (height) x 576 mm (depth)
- Outside: 1250 mm (width) x 1925 mm (height) x 890 mm (depth)
-
Light data per light cassette:
- Daylight lamps: 38 W/m², 13000 lx
- Fluora growth lamps: 36 W/m², 10500 lx
- Arabidopsis lamps: 43 W/m², 14000 lx
-
Electrical data:
- Nominal voltage: 200-240 V
- Mains frequency: 50/60 Hz
- Rated power: 2.7 kW
- Device fuse: 16 A
The KBW 720 is equipped with a BINDER controller that provides a clear display of all important information. The menu navigation is uncomplicated and enables simple operation. Further equipment features include:
- Electronic chart recorder
- Various graphical display options for the process parameters
- VENTAIR™ air jacket system (depending on model)
- APT.line™ preheating chamber technology
- Natural convection
- Adjustable exhaust air flap
- Timer function
- Integrated temperature selector limiter with visual alarm
- Door lock
- BINDER test certificate
Certifications and standards
The BINDER KBW 720 growth chamber meets the requirements of the BINDER factory standard and is based on DIN 12880:2007. These standards guarantee the reliability and precision of the unit, which is essential for carrying out reproducible experiments.
Application examples
The KBW 720 is ideal for a wide range of applications in laboratory and medical facilities:
-
Cultivation of microorganisms: create optimal conditions for the growth and multiplication of bacteria and fungi.
-
Cell cultures: Observe and analyze the behavior and growth of cells under controlled conditions.
-
Animal cultures: Support the development and growth of insect and small animal cultures.
-
Incubation and development processes: Use the cabinet for the incubation and development of eggs and larvae.
FAQs
How can I set the temperature in the BINDER KBW 720 growth chamber?
The temperature can be set using the integrated BINDER controller. This provides precise control and a clear display of the current temperature.
Which lamps are integrated in the KBW 720?
The KBW 720 is equipped with daylight lamps, Fluora growth lamps and Arabidopsis lamps. These offer different light intensities and spectra that are suitable for different applications.
How many shelves can I use in the KBW 720?
The KBW 720 offers space for up to 12 shelves, with 3 shelves included as standard.
Is the KBW 720 suitable for continuous operation?
Yes, the KBW 720 is designed for continuous operation and offers reliable performance over long periods of time.
Conclusion
The BINDER KBW 720 growth chamber is a powerful and versatile unit that no laboratory should be without. With its precise temperature and humidity control, extensive features and ease of use, it offers optimal conditions for a wide range of applications. Comparisons such as with the BINDER KBF P 720 constant climate chamber show that the KBW 720 is an excellent choice for demanding laboratory work.
Load max.: |
101 - 150 kg |
Width outside (incubators,...): |
1250 mm |
Width inside (incubators,...): |
970 mm |
Maximum temperature (heat): |
+ 70 °C |
Height outside (incubators,...): |
1925 mm |
Height inside (incubators,...): |
1250 mm |
Interior volume: |
601 - 700 l |
Cabinet type: |
climate chamber, growth chamber |
Depth outside (incubators,...): |
890 mm |
Depth inside (incubators,...): |
576 mm |
Lowest temperature (cooling): |
0 °C |
Shipping weight: |
407,00 kg
|
Item weight: |
377,00 kg
|
BINDER Growth chamber |
KBW 720-230V1) |
Performance Data Temperature |
|
Temperature range without illumination cassettes [°C] |
0...70 |
Temperature range with 100% illumination [°C] |
5...60 |
Temperature variation with 100% illumination [± K] |
1.0 |
Temperature variation without illumination cassettes [± K] |
0.5 |
Temperature fluctuation without illumination cassettes [± K] |
0.1 |
Temperature fluctuation with 100% illumination [± K] |
0.1 |
Lightdata per Illumination Cassettes2) |
|
Daylight tubes [W/m2] |
38 |
Daylight tubes [lx] |
13000 |
Fluora growth lamps [W/m2] |
36 |
Fluora growth lamps [lx] |
10500 |
Arabidopsis lamps [W/m2] |
43 |
Arabidopsis lamps [lx] |
14000 |
Electrical data |
|
Rated Voltage [V] |
200...240 |
Power frequency [Hz] |
50/60 |
Nominal power [kW] |
2.7 |
Unit fuse [A] |
16 |
Phase (Nominal voltage) [ph] |
1 |
Outer dimensions not incl. fittings and connections |
|
Width net [mm] |
1250 |
Height net [mm] |
1925 |
Depth net [mm] |
890 |
Wall clearance back [mm] |
100 |
Wall clearance sidewise [mm] |
100 |
Doors |
|
Inner doors |
2 |
Unit doors |
2 |
Internal Dimensions |
|
Width [mm] |
970 |
Height [mm] |
1250 |
Depth [mm] |
576 |
Measures |
|
Interior volume [L] |
698 |
Net weight of the unit (empty) [kg] |
377 |
permitted load [kg] |
150 |
Load per rack [kg] |
45 |
Environment-specific data |
|
Energy consumption at 37 °C and with illumination [Wh/h] |
880 |
Sound-pressure level [dB(A)] |
59 |
Fixtures |
|
Number of shelves (std./max.) |
3/12 |
Number of illumination cassettes (std./max.) |
3/3 |
1) All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
2) Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.
Toller Kundenservice und schnelle Beantwortung aller Fragen. Wir benötigten eine komplette Laborausstattung für unseren neuen Standort. Das Team von MedSolut hat alles zu unserer Zufriedenheit geregelt.